beautypg.com

Calibration frequency, Other quality assurance procedures – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 153

background image

Model 9110E Instruction Manual

EPA Protocol Calibration

M9110E Rev B

139

NO

2

CONVERTER EFFICIENCY

if the NO

X

reading should drop to less than 96% of its starting value during this step, it

indicates the NO

2

converter is in need of troubleshooting or replacement. See Chapter 11

for further details.

While maintaining all other conditions, adjust the ozone generator to obtain several other

concentrations of NO

2

evenly spaced between the 80% URL point and the zero point.

Record the NO, NO

X

, and NO

2

concentrations for each additional point.

Calculate the resulting NO

2

concentrations as follows:

F

]

NO

[

*

F

+

]

[NO

-

]

[NO

=

]

NO

[

T

IMP

2

NO

REM

ORIG

OUT

2

Eq 8-9

Where [NO]

ORIG

is the NO concentration before the GPT ozone is turned on, and [NO]

REM

is

the NO remaining after GPT.

Plot the NO

2

concentration output by the instrument on the y-axis against the generated

NO

2

[NO

2

]

OUT

on the x-axis. The plot should be a straight line within the ± 2% linearity

criteria given for the NO

x

and NO channels. If the plot is not linear the most likely cause is

that the converter needs replacing. See Section 7.1.7 on NO

2

converter efficiency.

8.4. Calibration Frequency

To ensure accurate measurements of the NO, NO

X

, and NO

2

concentrations, calibrate the

analyzer at the time of installation, and re-calibrate it:

• No later than three months after the most recent calibration or performance audit

which indicated analyzer calibration to be acceptable.

• An interruption of more than a few days in analyzer operation.
• Any repairs which might affect its calibration.
• Physical relocation of the analyzer.
• Any other indication (including excessive zero or span drift) of possible significant

inaccuracy of the analyzer.

Following any of the activities listed above, the zero and span should be checked to

determine if a calibration is necessary. If the analyzer zero and span drifts exceed the

calibration limits in Section 12 of the Q.A. Handbook

6

, a calibration should be performed.

8.5. Other Quality Assurance Procedures

Precision is determined by a one-point check at least once every two weeks. Accuracy is
determined by a three-point audit once each quarter.

Essential to quality assurance are scheduled checks for verifying the operational status of
the monitoring system. The operator should visit the site at least once each week. Every

two weeks a Level 1 zero and span check must be made on the analyzer. Level 2 zero and

span checks should be conducted at a frequency desired by the user. Definitions of these

terms are given in Table 8-3.