Nnn.nnn.nnn.nnn – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 100
Operating Instructions
Model 9110E Instruction Manual
86
M9110E Rev B
Ethernet and is used to set configuration parameters that enable the Ethernet card to
communicate with your local area network. To perform this configuration, you will need to
get the following information from your network administrator:
• Gateway IP address, typically a string of numbers with four packets of 1 to 3
numbers each:
nnn.nnn.nnn.nnn
, for example:
192.168.76.1
.
• Instrument IP address, typically a string of numbers very similar to the Gateway IP
address, for example:
192.168.76.55
. Note that this instrument IP address must
be a static address and cannot be a dynamic IP address. DHCP and analyzer host
names are not supported at this time.
• Subnet Mask, typically a string of numbers such as
255.255.252.0
.
• The communications port number is set to 3000 by default and should not be
changed.
The following procedures assume that the Ethernet card is already installed in your
analyzer. If you are retro-fitting the analyzer with an Ethernet Option 63, please make sure
to enable the internet option as described in Section 6.9.6 and set the baud rate to 115000
as described in Section 6.9.7. Refer to the installation instructions of that option for more
information.
Once Ethernet is enabled for the COM2 port, the first step in configuring the Ethernet option
is to set the card from pass-through mode to configuration mode, so that TCP/IP parame-
ters can be edited.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
SETUP
ETHERNET STOPPED
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2
EXIT
INTERNET CONFIGURATION
GTWY IP SNET PORT STRT STOP EXIT
STOPPING ETHERNET
If the message ETHERNET NOT STOPPED
appears here, the Ethernet card either was
already in configuration mode or might not
be responding.