Model 9110E Instruction Manual

Operating Instructions

M9110E Rev B

87

Step two in the configuration process is to edit the TCP/IP parameters.

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM VARS DIAG

EXIT

SAMPLE RANGE = 500.0 PPB O3 =XXX.X

< TST TST > CAL

SETUP

SETUP X.X COMMUNICATIONS MENU

ID INET COM1 COM2

EXIT

INTERNET CONFIGURATION

GTWY IP SNET PORT STRT STOP EXIT

SETUP X.X

GATEWAY IP: [0] 00.000.000


INS DEL [0]

ENTR EXIT

Internet Configuration

Keypad Functions

KEY FUNCTION

Moves the cursor one

character to the left.

CH>

Moves the cursor one

character to the right.

INS

Inserts a character before the

cursor location.

DEL

Deletes a character at the

cursor location.

[0]

Press this key to cycle

through the range of

numerals and characters

available for insertion.

ENTR

Accepts the new setting and

returns to the previous menu.

EXIT

Ignores the new setting and

returns to the previous menu.

Some keys only appear as needed.

INTERNET CONFIGURATION

GTWY IP SNET PORT STRT STOP EXIT

SETUP X.X

INST IP: [0] 00.000.000


INS DEL [0]

ENTR EXIT

INTERNET CONFIGURATION

GTWY IP SNET PORT STRT STOP EXIT

SETUP X.X

SUBNET MASK: [0] 00.000.000


INS DEL [0]

ENTR EXIT

INTERNET CONFIGURATION

GTWY IP SNET PORT STRT STOP EXIT

Cursor location is indicated by brackets

The PORT number

needs to remain at

3000. Do not change

this setting!

Change the cursor value with this button