beautypg.com

Editing idas parameters – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 112

background image

Operating Instructions

Model 9110E Instruction Manual

98

M9110E Rev B

6.10.2.5. Editing iDAS Parameters

Data channels can be edited individually from the front panel without affecting other data

channels. However, when editing a data channel, such as during adding, deleting or editing
parameters, all data for that particular channel will be lost, because the iDAS can store only

data of one format (number of parameter columns etc.) for any given channel. In addition,

an iDAS configuration can only be uploaded remotely as an entire set of channels. Hence,

remote update of the iDAS will always delete all current channels and stored data.

To modify, add or delete a parameter, press SETUP-DAS-ENTR and then:


Edit Data Parameter Menu

Edit Data Channel Menu

SETUP X.X 0) CONC: ATIMER, 2, 900

PREV NEXT INS DEL EDIT PRNT EXIT

FROM THE DAS PASSWORD MENU

Exits to the main

Data Acquisition

menu

SETUP X.X NAME:CONC

SET> EDIT PRINT

EXIT

SETUP X.X PARAMETERS:2

EDIT PRINT

EXIT

SETUP X.X 0) PARAM=DETREP, MODE=INST

PREV NEXT INS DEL EDIT EXIT

Press SET> key until…

SETUP X.X EDIT PARAMS (DELETE DATA)

YES NO

NO returns to

the previous

menu and

retains all data.

Moves the

display between

available

Parameters

Inserts a new Parameter

before the currently

displayed Parameter

Deletes the Parameter

currently displayed.

Use to configure

the functions for

this Parameter.

Exits to the main

Data Acquisition

menu

YES will delete

all data in that

entire channel.