beautypg.com

Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 135

background image

Model 9110E Instruction Manual

Calibration Procedures

M9110E Rev B

121

permeation tube and through the NO

2

converter (Figure 3-11). It also programs the

analyzer to use the NO output from the NO

2

converter to calibrate the span value of both

NO and NO

X

.

NOTE

The IZS with NO

2

permeation tube has a limited accuracy of about ±5%. Whereas

this may be sufficient for informal calibration checks, this method of calibration is

not approved by the US EPA as a calibration source.

This feature should only be enabled when a span calibration or calibration check is
performed. While CAL-ON-NO2 is enabled, the NO

2

concentration will always be reported

as zero. This is because the gas is continuously routed through the NO

2

converter and the

analyzer’s firmware simulates calibration with NO gas. Since the instrument sees the same

concentration of NO during both NO and NO

X

cycles, it reports an NO

2

concentration of zero.

Table 7-1: IZS Option Valve States with CAL_ON_NO

2

Turned ON

Valve

Condition

Valve Port Connections

Sample/Cal

Open to zero/span valve

1 Æ 2

Zero/Span

Open to SPAN GAS inlet

1 Æ 2

NO/NOx Valve

Open to NO2 converter

1 Æ 2

AutoZero Valve

Cycles normally

N/A

Performing a manual calibration of the analyzer with an IZS option installed is similar to the

method described in Section 7.4, except that both zero air and span gas are supplied to the

analyzer from ambient air through the zero gas inlet.

1. Connect the sources of zero air and span gas as shown in Figure 7-3.

Figure 7-3: Pneumatic Connections for Manual Calibration Checks with IZS

Set the expected NO and NO

X

values according to Section 7.2.

Perform the zero point calibration: