beautypg.com

Performing a sample flow check – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 223

background image

Model 9110E Instruction Manual

Troubleshooting & Repair

M9110E Rev B

209

• Install a leak checker or a tank of gas (compressed, oil-free air or nitrogen) as

described above on the sample inlet at the rear panel.

• Disconnect the pump tubing on the outside rear panel and cap the pump port. If IZS

or zero/span valves are installed, disconnect the tubing from the zero and span gas

ports and plug them (Figure 3-3). Cap the DFU particle filter on the Perma Pure

dryer (Figure 9-2).

• Pressurize the instrument with the leak checker or tank gas, allowing enough time to

fully pressurize the instrument through the critical flow orifice. Check each tube

connection (fittings, hose clamps) with soap bubble solution, looking for fine

bubbles. Once the fittings have been wetted with soap solution, do not re-apply

vacuum as it will draw soap solution into the instrument and contaminate it. Do not

exceed 15 psi pressure.

• If the instrument has the zero and span valve option, the normally closed ports on

each valve should also be separately checked. Connect the leak checker to the

normally closed ports and check with soap bubble solution.

• If the analyzer is equipped with an IZS Option Connect the leak checker to the Dry

Air inlet and check with soap bubble solution.

• Once the leak has been located and repaired, the leak-down rate of the indicated

pressure should be less than 1 in-Hg-A (0.4 psi) in 5 minutes after the pressure is

turned off.

• Clean surfaces from soap solution, re-connect the sample and pump lines and

replace the instrument cover. Restart the analyzer.

11.5.3. Performing a Sample Flow Check

CAUTION

Use a separate, calibrated flow meter capable of measuring flows between 0 and

1000 cm³/min to measure the gas flow rate though the analyzer. Do not use the

built in flow measurement viewable from the front panel of the instrument. This

value is only calculated, not measured.

Sample flow checks are useful for monitoring the actual flow of the instrument, as the front

panel display shows only a calculated value. A decreasing, actual sample flow may point to

slowly clogging pneumatic paths, most likely critical flow orifices or sintered filters. To
perform a sample flow check:

• Disconnect the sample inlet tubing from the rear panel SAMPLE port shown in Figure

3-3.

• Attach the outlet port of a flow meter to the sample inlet port on the rear panel.

Ensure that the inlet to the flow meter is at atmospheric pressure.

• The sample flow measured with the external flow meter should be 500 cm³/min ±

10%. If a combined sample/ozone air Perma Pure dryer is installed (optional equip-

ment), the flow will be 640 cm³/min ± 10% (500 cm³/min for the sample and 80

cm³/min for the ozone generator supply air and 60 cm³/min for the purge flow).

• Low flows indicate blockage somewhere in the pneumatic pathway.