Calibration quality analysis, The values for these – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 142
Calibration Procedures
Model 9110E Instruction Manual
128
M9110E Rev B
SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30
PREV NEXT MODE SET
EXIT
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG ACAL DAS RNGE PASS CLK MORE EXIT
SETUP X.X SEQ 1) DISABLED
NEXT MODE
EXIT
SETUP X.X SEQ 2) DISABLED
PREV NEXT MODE
EXIT
SETUP X.X MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X MODE: ZERO
PREV NEXT
ENTR EXIT
SETUP X.X MODE: ZERO–SPAN
PREV NEXT
ENTR EXIT
SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT MODE SET
EXIT
SETUP X.X STARTING DATE: 01–JAN–02
EXIT
SETUP X.X STARTING DATE: 01–JAN–02
0 4 SEP 0 3
ENTR EXIT
Toggle
keys to
set day,
month &
year: DD-
MON-Y
Y
SETUP X.X STARTING DATE: 04–SEP–03
EXIT
Default
value
is ON
SETUP C.4 STARTING DATE: 04–SEP–03
EXIT
Toggle keys to
set time:
HH:MM. This is
a 24 hr clock.
PM hours are
13-24.
Example: 2:15
PM = 14:15
Toggle keys
to set
number of
days
between
procedures
(1-367)
Toggle keys
to set
delay time for
each iteration
of the
sequence:
HH:MM
(0 – 24:00)
SETUP C.4 DELTA DAYS: 1
EXIT
SETUP C.4 DELTA DAYS: 1
0 0 2
ENTR EXIT
SETUP C.4 DELTA DAYS:2
EXIT
SETUP C.4 DELTA TIME00:00
EXIT
SETUP C.4 DELTA TIME: 00:00
0 0 :3 0
ENTR EXIT
SETUP C.4 DELTA TIEM:00:30
EXIT
SETUP C.4 STARTING TIME:14:15
EXIT
EXIT returns
to the SETUP
Menu
SETUP C.4 DURATION:15.0 MINUTES
EXIT
SETUP C.4 DURATION 15.0MINUTES
3 0 .0
ENTR EXIT
SETUP C.4 DURATION:30.0 MINUTES
EXIT
Toggle keys
to set
duration for
each
iteration of
the
sequence:
Set in
Decimal
minutes
from
0.1 – 60.0
SETUP C.4 CALIBRATE: OFF
EXIT
SETUP C.4 CALIBRATE: OFF
ON
ENTR EXIT
SETUP C.4 CALIBRATE: ON
EXIT
Toggle key
between
Off and
ON
Sequence #
Delta Time
Mode Delta Days
SETUP C.4 STARTING TIME:00:00
1 4 : 1 5
ENTR EXIT
SETUP X.X TIMER ENABLE: ON
SET> EDIT
EXIT
SETUP C.4 STARTING TIME:00:00
EXIT
7.9. Calibration Quality Analysis
After completing one of the calibration procedures described above, it is important to
evaluate the analyzer’s calibration SLOPE and OFFSET parameters. These values describe
the linear response curve of the analyzer, separately for NO and NO
X
. The values for these