First calibration, Basic calibration procedure – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 41

background image

Model 9110E Instruction Manual

Getting Started

M9110E Rev B

27

3.6. First Calibration

3.6.1. Basic Calibration Procedure

The initial calibration should be carried out using the 500 ppb range, a span gas with about

400 ppb NO and with the unit set for SINGLE range, which will enable you to compare your

calibration results to the factory calibration as listed on the Final Test and Validation Data

Sheet

. The following three-step procedure assumes that the instrument does not have any

of the available zero/span (Z/S) valve options installed. Chapter 7 contains instructions for

calibrating instruments with Z/S valve options, Chapter 8 for EPA protocol calibration.

Set the analog output range of the M9110E:

SETUP X.X RANGE CONTROL MENU

MODE

SET UNIT

EXIT

Press this button to set

the analyzer for

SNGL

DUAL or IND ranges

SETUP X.X

RANGE: 500.0 CONC

0 0 5 0 0 .0 ENTR EXIT

To change the value of the reporting
range span, change the numbers by

pressing the keys under each digit.

Press this button to select the

concentration units of measure:

PPB, PPM, UGM, MGM

EXIT ignores the new setting and returns
to the RANGE CONTROL MENU. ENTR

accepts the new setting and returns to the

RANGE CONTROL MENU.

SETUP X.X

CFG DAS RNGE PASS CLK MORE

EXIT

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL

SETUP

SETUP X.X

RANGE: 500.0 Conc

0 0 0 5 0 .0 ENTR EXIT