Model 9110E Instruction Manual

Troubleshooting & Repair

M9110E Rev B

213

Table 11-6: Analog Output Test Function - Nominal Values

Full scale Output Voltage

100mV 1V

5V 10V

Step

%

Nominal Output Voltage

1 0

0

mV

0 0 0

2 20

20

mV

0.2 1 2

3 40

40

mV

0.4 2 4

4 60

60

mV

0.6 3 6

5 80

80

mV

0.8 4 8

6 100

100

mV

1.0 5 10

11.5.9.3. Status Outputs

The procedure below can be used to test the Status outputs.

• Connect a cable between the “-“ pin and the “V” pin on the status output connector.

Table 11-7: Status Outputs Check

PIN #

Status

1 SYSTEM

OK

2 CONC

VALID

3 HIGH

RANGE

4 ZERO

CAL

5 SPAN

CAL

6 DIAG

MODE

7 SPARE
8 SPARE

• Connect a 1000 Ω resistor between the +5 V and the pin for the status output that is

being tested.

• Connect a voltmeter between the “-“ pin and the pin of the output being tested

(Table 11-7).

• Under the DIAG / SIGNAL I/O menu (Section 11.1.3), scroll through the inputs

and outputs until you get to the output in question. Alternately turn on and off the

output noting the voltage on the voltmeter, it should vary between 0 volts for ON
and 5 volts for OFF.

11.5.9.4. Control Inputs

The control input bits can be tested by the following procedure:

• Connect a jumper from the +5 V pin on the STATUS connector to the +5 V on the

CONTROL IN connector.