Status outputs, Control inputs – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 227
Model 9110E Instruction Manual
Troubleshooting & Repair
M9110E Rev B
213
Table 11-6: Analog Output Test Function - Nominal Values
Full scale Output Voltage
100mV 1V
5V 10V
Step
%
Nominal Output Voltage
1 0
0
mV
0 0 0
2 20
20
mV
0.2 1 2
3 40
40
mV
0.4 2 4
4 60
60
mV
0.6 3 6
5 80
80
mV
0.8 4 8
6 100
100
mV
1.0 5 10
11.5.9.3. Status Outputs
The procedure below can be used to test the Status outputs.
• Connect a cable between the “-“ pin and the “V” pin on the status output connector.
Table 11-7: Status Outputs Check
PIN #
Status
1 SYSTEM
OK
2 CONC
VALID
3 HIGH
RANGE
4 ZERO
CAL
5 SPAN
CAL
6 DIAG
MODE
7 SPARE
8 SPARE
• Connect a 1000 Ω resistor between the +5 V and the pin for the status output that is
being tested.
• Connect a voltmeter between the “-“ pin and the pin of the output being tested
(Table 11-7).
• Under the DIAG / SIGNAL I/O menu (Section 11.1.3), scroll through the inputs
and outputs until you get to the output in question. Alternately turn on and off the
output noting the voltage on the voltmeter, it should vary between 0 volts for ON
and 5 volts for OFF.
11.5.9.4. Control Inputs
The control input bits can be tested by the following procedure:
• Connect a jumper from the +5 V pin on the STATUS connector to the +5 V on the
CONTROL IN connector.