Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 253

background image

Model 9110E Instruction Manual

APPENDIX A

044100110 Rev B

239

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

NO2_SPAN2 Conc.

400 4-20000

Target

NO

2

concentration during

converter efficiency calibration of
range 2.

NOX_SLOPE2 PPB/mV

3,5

PPM/mV

4

1 0.25-4

NO

X

slope for range 2.

NOX_OFFSET2 mV

0

-10000-

10000

NO

X

offset for range 2.

NO_SLOPE2 PPB/mV

3,5

PPM/mV

4

1

0.25-4

NO slope for range 2.

NO_OFFSET2 mV 0

-10000-
10000

NO offset for range 2.

RANGE_MODE —

SNGL SNGL,

IND,

AUTO

Range control mode. Enclose value
in double quotes (") when setting

from the RS-232 interface.

PHYS_RANGE1 PPM 2

0.1-2500

Low

pre-amp

range.

PHYS_RANGE2

PPM

20

0.1-2500 High pre-amp range.

CONC_RANGE1 Conc. 500

10-
20000

D/A concentration range 1 or
range for NO

X

.

CONC_RANGE2

1

Conc.

500

10-

20000

D/A concentration range 2 or

range for NO.

CONC_RANGE3

1

Conc.

500

10-
20000

D/A concentration range 3 or
range for NO

2

.

RCELL_SET ºC 50

Warnings:

45-55

30-70

Reaction cell temperature set point
and warning limits.

MANIFOLD_SET

1

ºC

50

Warnings:

45-55

30-70

Manifold temperature set point and
warning limits (if installed).

CONV_TYPE — MOLY

NONE,

MOLY,
CONV,

O3KL

Converter type. “CONV” is mini-

hicon, O3KL is O

3

killer only (no

converter installed). Enclose value

in double quotes (") when setting

from the RS-232 interface.

CONV_SET ºC 315

Warnings:

305-325

0-800

Converter temperature set point
and warning limits for moly

converter.

CONV_TEMP_TRIG

Cycles

10

0-100

Number of converter temperature

errors required to trigger warning.

BOX_SET ºC

30

Warnings:

7-48

0-70

Nominal box temperature set point
and warning limits.

PMT_SET ºC

7

Warnings:

5-12

0-40

PMT temperature warning limits.

Set point is not used.

SFLOW_SET cc/m

500

Warnings:
350–600

0–1000

Sample flow warning limits. Set

point is not used.

SAMP_FLOW_SLOPE —

1

0.001-
100

Slope term to correct sample flow
rate.

OFLOW_SET

cc/m

80

0–500

Ozone flow warning limits. Set
point is not used.

OZONE_FLOW_SLOPE —

1

0.001-
100

Slope term to correct ozone flow
rate.