Model 9110E Instruction Manual

68

M9110E Rev B

6.7.3.3. Manual Analog Output Calibration

For highest accuracy, the voltages of the analog outputs can be manually calibrated. Note

that outputs configured for 0.1V full scale should always be calibrated manually. Calibration
is done through the instrument software with a voltmeter connected across the output

terminals (Figure 0-3). Adjustments are made using the front panel keys by setting the

zero-point first and then the span-point (Table 0-11). The software allows this adjustment

to be made in 100, 10 or 1 count increments.

Table 0-11: Voltage Tolerances for Analog Output Calibration

Full Scale

Zero Tolerance

Span Voltage

Span Tolerance

0.1 VDC

±0.0005V

90 mV

±0.001V

1 VDC

±0.001V

900 mV

±0.001V

5 VDC

±0.002V

4500 mV

±0.003V

10 VDC

±0.004V

4500 mV

±0.006V

V

+DC Gnd

V OUT +

V OUT -

V IN +

V IN -

Recording

Device

ANALYZER

See Table 3-1 for

pin assignments

of Analog Out

connector on the

rear panel

Figure 0-2: Setup for Calibrating Analog Outputs

To make these adjustments, the AOUT auto-calibration feature must be turned off

(Section 6.7.3). Then press SETUP-MORE-DIAG-ENTR-NEXT-NEXT and the following keys: