Operating Instructions

Model 9110E Instruction Manual

66

M9110E Rev B

6.7.3.2. Analog Output Calibration Mode

The analog outputs can be calibrated automatically or manually. In its default mode, the

instrument is configured for automatic calibration of all channels. Manual calibration should
be used for the 0.1V range or in cases where the outputs must be closely matched to the

characteristics of the recording device. Outputs configured for automatic calibration can be

calibrated as a group or individually. Analog calibration needs to be carried out on first

startup of the analyzer (performed in the factory as part of the configuration process) or

whenever re-calibration is required.

To calibrate the outputs as a group press SETUP-MORE-DIAG-ENTR-NEXT-NEXT and

then the following keys:

DIAG AIO AUTO CALIBRATING CONC_OUT_1

AUTO CALIBRATING CONC_OUT_2

AUTO CALIBRATING TEST_OUTPUT

Exit at any time

to return to the

main DIAG

menu

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED: NO

< SET SET> CAL

EXIT

If AutoCal has been

turned off for any

channel, the message

for that channel will be

similar to:

NOT AUTO CAL

CONC_OUT_1

Exit to return to

the I/O

configuration

menu

If any of the

channels

have not been

calibrated this

message will

read NO.

DIAG AIO AOUTS CALIBRATED: YES

< SET SET> CAL

EXIT

To automatically calibrate an analog channel, press SETUP-MORE-DIAG-ENTR-NEXT-NEXT
and then the following keys: