Instrument Maintenance

Model 9110E Instruction Manual

150

M9110E Rev B

9.3.3. Changing the Ozone Filter Chemical

The O

3

filter is located between the O

3

generator and the reaction cell and cleans the O

3

stream from solid and liquid contaminants that are created inside the O

3

generator. The

content of the filter cartridge needs periodical exchange according to Table 9-1, a rebuild kit
is available from the factory. Appendix B of this manual lists the part numbers of the ozone

filter rebuild kit. Some older models 9110E contain a Teflon cartridge, which is shown in
the inset. Use a special tool FT267 for the red fittings of those cartridges, otherwise follow

these instructions:

• To change the filter chemical, follow these steps:
• Turn of power to the analyzer and pump. Remove the analyzer cover and locate the

O

3

filter in the front of the analyzer next to the O

3

generator.

• Use a 7/16” wrench to remove both pieces of 1/8” tubing from the filter.

Figure 9-3: 04419 Ozone Filter Unit

• Remove the two 3” mounting screws with a Phillips screw driver and take out the

filter cartridge.

• Using a 9/16” wrench, remove both fittings from the cartridge.