Model 9110e, Insufficient gas flow can build up no – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 30

background image

Getting Started

Model 9110E Instruction Manual

16

M9110E Rev B


External Zero

Air Scrubber

Filter

Pump

MODEL 701

Zero Air

Generator

Source of

SAMPLE Gas

VENT if input is pressurized

Zero/Span Valves – Option 50

MODEL 700

Gas Dilution Calibrator

(with O

3

generator option)

VENT


MODEL

9110E

Sample

Zero Air

Exhaust

Span

Source of

SAMPLE Gas

VENT if input is pressurized

Internal Zero/Span Option (IZS) – Option 51

Pump


MODEL

9110E

Sample

Zero Air

Exhaust

Span

External Zero

Air Scrubber

Particulate

Filter

Calibrated NO gas

(high concentration)

Figure 3-6: Basic Pneumatic Connections for Units With Valve Options

Once the appropriate pneumatic connections have been made, check all pneumatic fittings
for leaks using the procedures defined in Section 11.5.

WARNING

Gas flow must be maintained at all times for units with IZS Options installed. The

IZS option includes a permeation tube which emits NO

2

. Insufficient gas flow can

build up NO

2

to levels that will damage the instrument. Remove the permeation

device when taking the analyzer out of operation.