Model 9110E Instruction Manual

Calibration Procedures

M9110E Rev B

117

7.2.1. Connect Zero Air and Span Gases to the Analyzer

Pump

MODEL

9110E

Sample

Zero Air

Exhaust

Span

MODEL 700 Gas

Dilution

Calibrator

(with O

3

generator

option)

MODEL 701

Zero Air

Generator

Vent excess gas

Calibrated NO gas

(high concentration)

Source of

sample gas

removed

during

calibration

Valve

OR

Needle valve to control flow

MODEL

9110E

Sample

Zero Air

Exhaust

Span

VENT

Calibrated NO gas

(At span gas concentration)

Source of

SAMPLE Gas

removed

during

calibration

Pump

MODEL 701

Zero Air

Generator

Figure 7-2: Setup for Manual Calibration without Z/S valve or IZS Option

7.2.2. Set Expected

NO and NO

X

S

pan Gas Concentrations

SAMPLE RANGE = 500.0 PPB NOX= X.X

< TST TST > CAL

SETUP

This sequence causes the

analyzer to prompt for the

expected NO

x

span

concentration.

The NO

X

& NO span concentration

values automatically default to

400.0 Conc.

To change this value to the

concentration of the span gas, enter

the number by pressing the key

under each digit until the value is

set.

EXIT ignores the new setting and

returns to the

CONCENTRATION MENU.

ENTR accepts the new setting and

returns to the

CONCENTRATION MENU..

M-P CAL CONCENTRATION MENU

NOX NO CONV

EXIT

M-P CAL NOX SPAN CONC: 400.0 Conc

0 0 4 0 0 .0 ENTR EXIT

M-P CAL CONCENTRATION MENU

NOX NO CONV

EXIT

M-P CAL NO SPAN CONC: 450.0 Conc

0 0 4 5 0 .0 ENTR EXIT

M-P CAL RANGE = 500.0 PPB NO = X.X

< TST TST > ZERO CONC

EXIT