beautypg.com

Using the diagnostic signal i/o function – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 209

background image

Model 9110E Instruction Manual

Troubleshooting & Repair

M9110E Rev B

195

measurements recorded on the factory data sheet may also indicate a failure or a

maintenance item. A problem report worksheet has been provided in Appendix C (TAI part

number 04503) to assist in recording the value of these test functions. The following table

contains some of the more common causes for these values to be out of range.

Table 11-1: Test Functions - Possible Causes for Out-Of-Range Values

Test Function

Indicated Failure(s)

NOx STB

Unstable concentrations; leaks

SAMPLE Fl

Leaks; clogged critical flow orifice

OZONE FL

Leaks; clogged critical flow orifice

PMT

Calibration off; HVPS problem; no flow (leaks)

NORM PMT

AutoZero too high

AZERO

Leaks; malfunctioning NONOx or AutoZero valve; O

3

air filter cartridge

exhausted

HVPS

HVPS broken; calibration off; preamp board circuit problems

RCELL TEMP

Malfunctioning heater; relay board communication (I

2

C bus); relay burnt out

BOX TEMP

Environment out of temperature operating range; broken thermistor

PMT TEMP

TEC cooling circuit broken; relay board communication (I

2

C bus); 12 V power

supply

IZS TEMP (option)

Malfunctioning heater; relay board communication (I

2

C bus); relay burnt out

MOLY TEMP

Malfunctioning heater; disconnected or broken thermocouple; relay board

communication (I

2

Z bus); relay burnt out; incorrect AC voltage configuration

RCEL (pressure)

Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices

SAMP (pressure)

Leak; malfunctioning valve; malfunctioning pump; clogged flow orifices;
sample inlet overpressure;

NOX SLOPE

HVPS out of range; low-level (hardware) calibration needs adjustment; span
gas concentration incorrect; leaks

NOX OFF

Incorrect span gas concentration; low-level calibration off

NO SLOPE

HVPS out of range; low-level calibration off; span gas concentration incorrect;
leaks

NO OFFS

Incorrect span gas concentration; low-level calibration off

Time of Day

Internal clock drifting; move across time zones; daylight savings time?

11.1.3. Using the Diagnostic Signal I/O Function

The signal I/O parameters found under the diagnostics (DIAG) menu combined with a

thorough understanding of the instrument’s theory of operation (Chapter 10) are useful for

troubleshooting in three ways:

• The technician can view the raw, unprocessed signal level of the analyzer’s critical

inputs and outputs.

• All of the components and functions that are normally under instrument control can

be manually changed.

• Analog and digital output signals can be manually controlled.

This allows to systematically observe the effect of these functions on the operation of the

analyzer. Figure 11-2 shows an example of how to use the signal I/O menu to view the raw