beautypg.com

Independent range mode, Calibration with remote contact closures – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 139

background image

Model 9110E Instruction Manual

Calibration Procedures

M9110E Rev B

125

SAMPLE RANGE = 500.0 PPB NOX=X.XX

< TST TST > CAL CALZ CALS

SETUP

ZERO CAL M

RANGE = 500.0 PPB X=X.XX


< TST TST > ZERO CONC

EXIT

SAMPLE

RANGE TO CAL: LOW


LOW HIGH

ENTR SETUP

SAMPLE

RANGE TO CAL: HIGH


LOW HIGH

ENTR SETUP

…continue Calibration

as per standard

procedure

Analyzer enters ZERO

CAL mode

Once this selection is made, the calibration/check procedure continues as described in

Sections 7.2 through 7.6. The other range may be calibrated by starting over from the main

SAMPLE display.

7.7.2. Independent Range Mode

Independent range mode is described in Section 6.5.4. When selecting span gas concentra-
tions for the NO/NO

X

or NO

2

dynamic calibrations, use concentrations around 80% of the

highest reporting range for all gases. This will produce the most accurate calibration.

The span calibration routine is only required to be performed once, unlike AUTO range

mode, which requires separate HIGH and LOW range calibrations. If the calibration data

are obtained from the RS-232 port or from the front panel display, no special changes are
necessary as the IND range mode only affects the scaling of the analog outputs.

7.7.3. Calibration With Remote Contact Closures

Contact closures for controlling calibration and calibration checks are located on the rear

panel CONTROL IN connector. Instructions for setup and use of these contacts can be

found in Section 6.8.2.

When the appropriate contacts are closed for at least 5 seconds, the instrument switches

into zero, low span or high span mode and internal zero/span valves (if installed) will be
automatically switched to the appropriate configuration. The remote calibration contact

closures may be activated in any order. It is recommended that contact closures remain
closed for at least 10 minutes to establish a reliable reading; the instrument will stay in the

selected mode for as long as the contacts remain closed.

If contact closures are used in conjunction with the analyzer’s AutoCal (Section 7.8) feature

and the AutoCal attribute CALIBRATE is enabled, the M9110E will not re-calibrate the

analyzer until the contact is opened. At this point, the new calibration values will be
recorded before the instrument returns to SAMPLE mode. If the AutoCal attribute

CALIBRATE is disabled, the instrument will return to SAMPLE mode, leaving the instru-
ment’s internal calibration variables unchanged.