beautypg.com

Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 143

background image

Model 9110E Instruction Manual

Calibration Procedures

M9110E Rev B

129

terms, both individually and relative to each other, indicate the quality of the calibration. To

perform this quality evaluation, you will need to record the values of the following test

functions (Section 6.2.1 or Appendix A-3), all of which are automatically stored in the iDAS

channel CALDAT for data analysis, documentation and archival.

NO OFFS
NO SLOPE
NOX OFFS
NOX SLOPE

Make sure that these parameters are within the limits listed in Table 7-5 and frequently

compare them to those values on the Final Test and Checkout Sheet that came attached to

your manual, which should not be significantly different. If they are, refer to the trouble-

shooting Chapter 11.

Table 7-5: Calibration Data Quality Evaluation

Function

Minimum Value

Optimum Value

Maximum Value

NOX SLOPE

-0.700

1.000

1.300

NO SLOPE

-0.700

1.000

1.300

NOX OFFS

-20.0 mV

0.0 mV

150.0 mV

NO OFFS

-20.0 mV

0.0 mV

150.0 mV

The default iDAS configuration records all calibration values in channel CALDAT as well as

all calibration check (zero and span) values in its internal memory. Up to 200 data points
are stored for up 4 years of data (on weekly calibration checks) and a lifetime history of

monthly calibrations. Review these data to see if the zero and span responses change over

time. These channels also store the STABIL figure (standard deviation of NO

X

concentration)

to evaluate if the analyzer response has properly leveled off during the calibration

procedure. Finally, the CALDAT channel also stores the converter efficiency for review and

documentation.