APPENDIX A
044100110 Rev B
243
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
RCELL_PRESS_SET In-Hg
6
Warnings:
0-15
0-100
Reaction cell pressure warning
limits. Set point is not used.
RCELL_CYCLE
Seconds
10
0.5-30
Reaction cell temperature control
cycle period.
RCELL_PROP
1/ºC
1
0-10
Reaction cell PID temperature
control proportional coefficient.
RCELL_INTEG
—
0.1
0-10
Reaction cell PID temperature
control integral coefficient.
RCELL_DERIV
—
0 (disabled)
0-10
Reaction cell PID temperature
control derivative coefficient.
MANIFOLD_CYCLE
1
Seconds
5
0.5-30 Manifold
temperature control cycle
period.
MANIFOLD_PROP
1
1/ºC
0.2
0-10
Manifold PID temperature control
proportional coefficient.
MANIFOLD_INTEG
1
—
0.1
0-10
Manifold PID temperature control
integral coefficient.
MANIFOLD_DERIV
1
—
0.5
0-10
Manifold PID temperature control
derivative coefficient.
SLOPE_CONST
—
8
0.1-100
Slope constant factor to keep
visible slope near 1.
I2C_RESET_ENABLE —
ON
OFF,
ON I
2
C bus automatic reset enable.
ALARM_TRIGGER
Cycles
10
1–100
Number of valve cycles to trigger
concentration alarm.
DISP_INTENSITY —
HIGH
HIGH,
MED,
LOW,
DIM
Front panel display intensity.
Enclose value in double quotes (")
when setting from the RS-232
interface.
SERIAL_NUMBER —
“00000000
”
Any
character
in the
allowed
character
set. Up
to 100
charac-
ters long.
Unique serial number for
instrument. Enclose value in
double quotes (") when setting
from the RS-232 interface.
Format: YYDDDNNNNS (year, Day
of Year, Number, Special)