Specifications, approvals and warranty, M9110e operating specifications – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 19
Model 9110E Instruction Manual
Specifications, Approvals and Warranty
M9110E Rev B
5
2. SPECIFICATIONS, APPROVALS AND
WARRANTY
2.1. M9110E Operating Specifications
Table 2-1: Model 9110E Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
Min: 0-50 ppb
Max: 0-20 ppm
Measurement Units
ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
Zero Noise
1
0.2
ppb
(RMS)
Span Noise
1
<0.5% of reading above 50 ppb
or 0.2 ppm, whichever is greater
Lower Detectable Limit
2
0.4
ppb
Zero Drift (24 hours)
<0.5 ppb (at constant temperature and voltage.)
Zero Drift (7 days)
1 ppb (at constant temperature and voltage.)
Span Drift (7 Days)
<0.5% of full Scale (at constant temperature and voltage.)
Linearity
1% of full scale
Precision
0.5% of reading
Temperature Coefficient
< 0.1% per °C
Voltage Coefficient
< 0.1% per V
Lag Time
1
20 s
Rise/Fall Time
1
95% in <60 s
Sample Flow Rate
500 cm
3
/min. ± 10%
Temperature Range
5 - 40 °C operating and EPA equivalency
Humidity Range
0-95% RH non-condensing
Dimensions H x W x D
18 cm x 43 cm x 61 cm (7" x 17" x 23.6")
Weight, Analyzer
18 kg (40 lbs)
Weight, Ext Pump Pack
7 kg (16 lbs)
AC Power Rating
100 V, 50/60 Hz (3.25A);
115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)
Power, Ext Pump
100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 - 240 V, 50/60 Hz (2.5 A)
Environmental
Installation category (over-voltage category) II; Pollution degree 2
Analog Outputs
4 outputs
Analog Output Ranges
All Outputs: 0.1 V, 1 V, 5 V or 10 V (user selectable)
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage (12 bit)
Status Outputs
8 Status outputs from opto-isolators, 7 defined, 1 spare
Control Inputs
6 Control inputs, 4 defined, 2 spare
Serial I/O
1 RS-232; 1 RS-485 or RS-232 (configurable)
Communication speed: 300 - 115200 baud (user selectable)
Certifications
USEPA: Reference Method Number RFNA 1194-099
CE: EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section
15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548
(w/A1 & A2; 97) Class A.
1
As defined by the USEPA.
2
Defined as twice the zero noise level by the USEPA.