Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 250
APPENDIX A
Model 9110E Instruction Manual
236
044100120 Rev B
APPENDIX A-2:
Setup Variables For Serial I/O, Revision C.8
Table A-1: M9110E Setup Variables, Revision C.8
Please note that variables listed below under “Setup Variables Available Through serial I/O”
should not be changed unless specifically instructed by Teledyne-API customer service or
engineering! Changing those variables may cause significant problems with analyzer
performance. Only those variables listed under “Setup Variables Available through Front
Panel Display And serial I/O” are those see on the front panel display and should be
changed by the user if necessary.
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
Setup Variables Available through Front Panel Display And serial I/O
DAS_HOLD_OFF
Minutes
15
0.5-20
Duration of DAS hold off period.
MEASURE_MODE —
NO-NOX,
NOX
8
NO
NO
X
NOX-NO
Gas measure mode: single gas or
switching. Enclose value in double
quotes (") when setting from the
RS-232 interface.
TPC_ENABLE — ON OFF,
ON
ON
enables temperature/ pressure
compensation; OFF disables it.
DYN_ZERO
—
OFF
ON, OFF
ON enables remote dynamic zero
calibration; OFF disables it.
DYN_SPAN
—
OFF
ON, OFF
ON enables remote dynamic span
calibration; OFF disables it.
IZS_SET
1
ºC
50
Warnings:
45-55
30-70
IZS temperature set point and
warning limits.
CONC_PRECISION —
AUTO
AUTO
0
1
2
3
4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
Enclose value in double quotes (")
when setting from the RS-232
interface.
CLOCK_ADJ Sec./Day
0 -60-60
Time-of-day
clock
speed
adjustment.
CAL_ON_NO2
—
OFF
ON, OFF
ON enables span calibration on
pure NO
2
; OFF disables it.
Setup Variables Available Through serial I/O
LANGUAGE_SELECT —
ENGL
ENGL,
SECD
EXTN
Selects the language to use for the
user interface. Enclose value in
double quotes (") when setting
from the RS-232 interface.
MAINT_TIMEOUT
Hours
2
0.1-100
Time until automatically switching
out of software-controlled
maintenance mode.
BXTEMP_TPC_GAIN
—
0.3
0-10
Box temperature compensation
attenuation factor.
RCTEMP_TPC_GAIN —
0
0-10 Reaction
cell
temperature
compensation attenuation factor.
RCPRESS_TPC_GAIN —
0
0-10
Reaction
cell
pressure
compensation attenuation factor.