APPENDIX A

Model 9110E Instruction Manual

236

044100120 Rev B

APPENDIX A-2:

Setup Variables For Serial I/O, Revision C.8

Table A-1: M9110E Setup Variables, Revision C.8

Please note that variables listed below under “Setup Variables Available Through serial I/O”

should not be changed unless specifically instructed by Teledyne-API customer service or

engineering! Changing those variables may cause significant problems with analyzer

performance. Only those variables listed under “Setup Variables Available through Front

Panel Display And serial I/O” are those see on the front panel display and should be
changed by the user if necessary.

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

Setup Variables Available through Front Panel Display And serial I/O

DAS_HOLD_OFF

Minutes

15

0.5-20

Duration of DAS hold off period.

MEASURE_MODE —

NO-NOX,
NOX

8

NO
NO

X

NOX-NO

Gas measure mode: single gas or
switching. Enclose value in double

quotes (") when setting from the
RS-232 interface.

TPC_ENABLE — ON OFF,

ON

ON

enables temperature/ pressure

compensation; OFF disables it.

DYN_ZERO

OFF

ON, OFF

ON enables remote dynamic zero
calibration; OFF disables it.

DYN_SPAN

OFF

ON, OFF

ON enables remote dynamic span
calibration; OFF disables it.

IZS_SET

1

ºC

50

Warnings:

45-55

30-70

IZS temperature set point and
warning limits.

CONC_PRECISION —

AUTO

AUTO

0
1

2

3
4

Number of digits to display to the

right of the decimal point for
concentrations on the display.

Enclose value in double quotes (")

when setting from the RS-232
interface.

CLOCK_ADJ Sec./Day

0 -60-60

Time-of-day

clock

speed

adjustment.

CAL_ON_NO2

OFF

ON, OFF

ON enables span calibration on
pure NO

2

; OFF disables it.

Setup Variables Available Through serial I/O

LANGUAGE_SELECT —

ENGL

ENGL,

SECD
EXTN

Selects the language to use for the

user interface. Enclose value in
double quotes (") when setting

from the RS-232 interface.

MAINT_TIMEOUT

Hours

2

0.1-100

Time until automatically switching
out of software-controlled

maintenance mode.

BXTEMP_TPC_GAIN

0.3

0-10

Box temperature compensation
attenuation factor.

RCTEMP_TPC_GAIN —

0

0-10 Reaction

cell

temperature

compensation attenuation factor.

RCPRESS_TPC_GAIN —

0

0-10

Reaction

cell

pressure

compensation attenuation factor.