Model 9110E Instruction Manual

Theory of Operation

M9110E Rev B

175

cm³/min), a warning message OZONE FLOW WARNING is displayed on the analyzer’s

front panel (Section 6.2.2) and the O

3

generator is turned off. As second warning, OZONE

GEN OFF, is displayed. This flow measurement is viewable through instrument’s front panel

display as the test function OZONE FL.

10.2.11. Dilution Manifold

Certain applications require to measure NO

X

in sample gases that do not contain any

oxygen. However, the molybdenum NO

2

converter requires a minimum amount of oxygen

to operate properly and to ensure constant conversion efficiency. For these special

applications, the M9110E analyzer may be equipped with a dilution manifold (Figure 10-11)

to provide the instrument with an internal sample stream that contains about 2.5% O

2

. This

manifold is mounted between converter housing and vacuum manifold on a small mounting

bracket. The manifold is equipped with two orifice holders that control the flow of the O

2

-

free sample gas and the bleeds in a small amount of zero air before the combined sample
stream goes to the NO/NO

X

valve for measurement. The zero air is produced by an external

zero air scrubber cartridge, mounted on the rear panel (Figure 3-3 and Figure 9-4).

Figure 10-11:

Dilution Manifold

Please inquire with Teledyne-API sales if the M9110E can be modified to fit your application.

10.3. Electronic Operation

Figure 10-12 shows a block diagram of the major electronic components of the M9110E.

The core of the analyzer is a microcomputer (CPU) that controls various internal processes,
interprets data, calculates data, and reports results using specialized firmware developed by

TAI. It communicates with the user, receives data from and issues commands to a variety