beautypg.com

Calibration procedures, Calibration preparations, Required equipment, supplies, and expendables – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 127: Zero air, Span gas standards

background image

Model 9110E Instruction Manual

Calibration Procedures

M9110E Rev B

113

7. CALIBRATION PROCEDURES

This chapter describes calibration procedures for the M9110E. All of the methods described
here can be initiated and controlled through the front panel or the COM ports.

NOTE

If you are using the M9110E for US-EPA controlled monitoring, see Chapter 8 for

information on the EPA calibration protocol.

7.1. Calibration Preparations

The calibration procedures in this section assume that the range mode, analog range and

units of measure have already been selected for the analyzer. If this has not been done,

please do so before continuing (Section 6.5 for instructions).

7.1.1. Required Equipment, Supplies, and Expendables

Calibration of the Model 9110E analyzer requires a certain amount of equipment and
supplies. These include, but are not limited to, the following:

• Zero-air source (defined in Section 3.3.2).
• Span gas source (defined in Section 3.3.1).
• Gas lines - all gas line materials should be stainless steel or Teflon-type (PTFE or

FEP). High concentration NO gas transported over long distances may require

stainless steel to avoid oxidation of NO with O

2

diffusing into the tubing.

• A recording device such as a strip-chart recorder and/or data logger (optional). For

electronic documentation, the internal data acquisition system can be used.

7.1.2. Zero Air

Zero air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all

components that might affect the analyzer’s readings. For NO

X

measuring devices, zero air

should be devoid of NO

X

and large amounts of CO

2

, NH

3

and water vapor. Water vapor and

moderate amounts of NH

3

can be removed using a sample gas conditioner (Section 5.9).

Devices such as the API Model 701 zero air generator that condition ambient air by drying

and removal of pollutants are available. We recommend this type of device for generating
zero air. Please contact our sales department for more information on this.

7.1.3. Span Gas Standards

As explained in detail in Section 3.3.1, span gas is specifically mixed to match the chemical
composition of the gas being measured at about 80% of the desired full measurement

range. For example, if the measurement range is 500 ppb, the span gas should have an NO
concentration of about 400 ppb.