Holdoff feature – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 118

background image

Operating Instructions

Model 9110E Instruction Manual

104

M9110E Rev B

To disable a data channel, press SETUP-DAS-ENTR and the following keys:


Edit Data Channel Menu

SETUP X.X 0) CONC: ATIMER, 2, 900

PREV NEXT INS DEL EDIT PRNT EXIT

FROM THE DAS PASSWORD MENU

Exits to the main

Data Acquisition

menu

SETUP X.X NAME:CONC

SET> EDIT PRINT

EXIT

SETUP X.X CHANNEL ENABLE:ON

EDIT PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X CHANNEL ENABLE:ON

OFF

ENTR EXIT

Toggle key to turn

channel ON or OFF

Press SET> key until…

6.10.2.12. HOLDOFF Feature

The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during
the DAS_HOLDOFF period enabled and specified in the VARS (Section 6.5). To enable or

disable the HOLDOFF for any one iDAS channel, press SETUP-DAS-ENTR and then:


Edit Data Channel Menu

SETUP X.X 0) CONC: ATIMER, 2, 900

PREV NEXT INS DEL EDIT PRNT EXIT

FROM THE DAS PASSWORD MENU

Exits to the main

Data Acquisition

menu

SETUP X.X NAME:CONC

SET> EDIT PRINT

EXIT

SETUP X.X CAL HOLD OFF:ON

SET> EDIT PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X CAL HOLD OFF:ON

ON

ENTR EXIT

Toggle key to turn

HOLDOFF ON or OFF

Press SET> key until…