Manual calibration – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 130

background image

Calibration Procedures

Model 9110E Instruction Manual

116

M9110E Rev B

Step three is to perform the converter efficiency calculation procedure:

M-P CAL CONVERTER EFFICIENCY MENU

NO

2

CAL SET

EXIT

M-P CAL CE FACTOR: 1.0000 GAIN

1. 0 0 0 0

ENTR EXIT

3x

EXIT to return to the

main SAMPLE display.

Wait until the

reading

stabilizes and

the ENTR

button is

displayed

ACTION:

Supply NO

2

span gas to the instrument.

NOTE:

The converter efficiency must be

between 0.9600 and 1.0200,

otherwise the NO

2

converter

needs to be replaced.

When ENTR is pressed, the ratio

of observed NO

2

concentration to

expected NO

2

concentration is

calculated and stored.

Continue from the

CONVERTER EFFICIENCY MENU

(step two above)

M-P CAL CONVERTER EFFICIENCY MENU

NO

2

CAL SET

EXIT

M-P CAL CONVERTER EFFICIENCY MENU

NO

2

CAL SET

EXIT

M-P CAL

CE FACTOR: 0.9852 GAIN

0. 9 8 5 2

ENTR EXIT

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > ENTR

EXIT

7.2. Manual Calibration

The following section describes the basic method for manually calibrating the Model 9110E
NO

X

analyzer.

A note on the difference between a calibration and calibration check: pressing the ENTR key

during the following procedures recalculates the stored values for OFFSET and SLOPE

(instrument response curve) and alters the instrument’s calibration. Alternatively, you can

check on the current values and response and then EXIT out of any menu before pressing

the final ENTR key to leave all settings intact (Section 7.3).