Model 9110E Instruction Manual
Operating Instructions
M9110E Rev B
55
Setting analog range limits to different values does not affect the instrument’s calibration.
To select the IND range mode, press the following keys:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE: SNGL
SNGL IND AUTO
ENTR EXIT
SETUP X.X
RANGE MODE: IND
SNGL
DUAL AUTO
ENTR EXIT
EXIT ignores the new setting and returns to the
RANGE CONTROL MENU.
ENTR accepts the new setting and returns to the
RANGE CONTROL MENU.
To set the range value for each independent range, press:
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X NO RANGE: 500.0 Conc
0 1 0 0 0 .0 ENTR EXIT
SETUP X.X NOX RANGE: 500.0 Conc
0 1 5 0 0 .0 ENTR EXIT
SETUP X.X NO2 RANGE: 500.0 Conc
0 0 5 0 0 .0 ENTR EXIT
Press the
number keys to
set range value
EXIT ignores the new
setting and returns to the
RANGE CONTROL
MENU.
ENTR accepts the new
setting and returns to the
RANGE CONTROL
MENU.
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT