Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 33
Model 9110E Instruction Manual
Getting Started
M9110E Rev B
19
If you wish to use the analyzer to remotely activate the zero and span calibration modes,
several digital control inputs are provided through a 10-pin connector labeled CONTROL IN
on the analyzer’s rear panel.
There are two methods for energizing the control inputs. The internal +5V available from
the pin labeled “+” is the most convenient method. However, if full isolation is required, an
external 5 VDC power supply should be used.
SPAN C
A
L
ZERO CA
L
CONTROL IN
Local Power Connections
External Power Connections
SPAN C
A
L
ZERO CA
L
CONTROL IN
-
+
5 VDC Power
Supply
A B C D E F U
+
A B C D E F U
+
Figure 3-9: Control Input Connector
Table 3-4: Control Input Signals
Input #
Status Definition
ON Condition
A
REMOTE ZERO CAL
The analyzer is placed in Zero Calibration mode. The mode
field of the display will read ZERO CAL R.
B
REMOTE SPAN CAL
The analyzer is placed in low span calibration mode as part
of performing a low span (midpoint) calibration. The mode
field of the display will read LO CAL R.
C SPARE
D SPARE
E SPARE
F SPARE
Digital Ground
The ground level from the analyzer’s internal DC power
supplies (same as chassis ground).
U
External Power input
Input pin for +5 VDC required to activate pins A - F.
+
5 VDC output
Internally generated 5V DC power. To activate inputs A - F,
place a jumper between this pin and the “U” pin. The
maximum amperage through this port is 300 mA (combined
with the analog output supply, if used).
If you wish to utilize one of the analyzer’s two serial ports, attach the serial cable that is
included with the analyzer to the COM1 serial port on the rear panel. Connect the other end
to a serial port on your computer and set the computer serial port to 115000 baud transfer