Model 9110E Instruction Manual

Operating Instructions

M9110E Rev B

77

The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel

labeled STATUS. The function of each pin is defined in Table 0-14.

CO

MMON

EMIT

T

E

RS

STATUS

1 2 3 4 5 6 7 8 D

+

SY

ST

E

M

O

K

HIGH RANGE

CO

NC VALI

D

ZERO CAL

SPA

N CAL

DIAG

N

O

S

T

IC MODE

LOW

SP

AN

GROUND

Figure 0-4: Status Output Connector

Table 0-14: Status Output Pin Assignments

Connector

Pin

Status Condition

(ON=conducting)

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

ON if concentration measurement is valid, OFF when invalid.

3

HIGH RANGE

ON if unit is in high range of any AUTO range mode.

4

ZERO CAL

ON whenever the instrument is in ZERO calibration mode.

5

SPAN CAL

ON whenever the instrument is in SPAN calibration mode.

6

DIAG MODE

ON whenever the instrument is in DIAGNOSTIC mode.

7

LOW RANGE

ON if unit is in low range of any AUTO range mode.

8

Unused.

D EMITTER

BUS

The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected

to the circuit ground of the receiving device.

+ DC

POWER

+ 5 VDC, 30 mA maximum (combined rating with Control
Inputs).

DIGITAL
GROUND

The ground from the analyzer’s internal, 5/±15 VDC power
supply.

6.8.2. Control Inputs

Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are

provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.

These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U”

pin is connected to the respective input pin.