Status, Control inputs – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 91
Model 9110E Instruction Manual
Operating Instructions
M9110E Rev B
77
The status outputs are accessed through a 12 pin connector on the analyzer’s rear panel
labeled STATUS. The function of each pin is defined in Table 0-14.
CO
MMON
EMIT
T
E
RS
STATUS
1 2 3 4 5 6 7 8 D
+
SY
ST
E
M
O
K
HIGH RANGE
CO
NC VALI
D
ZERO CAL
SPA
N CAL
DIAG
N
O
S
T
IC MODE
LOW
SP
AN
GROUND
Figure 0-4: Status Output Connector
Table 0-14: Status Output Pin Assignments
Connector
Pin
Status Condition
(ON=conducting)
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
ON if concentration measurement is valid, OFF when invalid.
3
HIGH RANGE
ON if unit is in high range of any AUTO range mode.
4
ZERO CAL
ON whenever the instrument is in ZERO calibration mode.
5
SPAN CAL
ON whenever the instrument is in SPAN calibration mode.
6
DIAG MODE
ON whenever the instrument is in DIAGNOSTIC mode.
7
LOW RANGE
ON if unit is in low range of any AUTO range mode.
8
Unused.
D EMITTER
BUS
The emitters of the transistors on pins 1-8 are bussed
together. For most applications, this pin should be connected
to the circuit ground of the receiving device.
+ DC
POWER
+ 5 VDC, 30 mA maximum (combined rating with Control
Inputs).
DIGITAL
GROUND
The ground from the analyzer’s internal, 5/±15 VDC power
supply.
6.8.2. Control Inputs
Control inputs allow the user to remotely initiate ZERO and SPAN calibration modes are
provided through a 10-pin connector labeled CONTROL IN on the analyzer’s rear panel.
These are opto-isolated, digital inputs that are activated when a 5 VDC signal from the “U”
pin is connected to the respective input pin.