Ozone generator override – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 89
Model 9110E Instruction Manual
Operating Instructions
M9110E Rev B
75
To activate the electrical test press the following keys.
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
DIAG
SIGNAL I / O
PREV NEXT JUMP
ENTR
EXIT
DIAG ELEC RANGE = 500.0 PPB NOX=X.X
EXIT
Press NEXT until…
DIAG
ELECTRICAL TEST
PREV NEXT
ENTR EXIT
Press TST until…
DIAG ELEC PMT = 1732 MV NOX=X.X
EXIT
While the electrical test is
activated, PMT should equal:
2000 mV ± 1000 mV
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR EXIT
6.7.7. Ozone Generator Override
This feature allows the user to manually turn the ozone generator off and on. This can be
done before disconnecting the generator, to prevent ozone from leaking out, or after a
system restart if the user does not want to wait for 30 minutes during warm-up time. Note
that this is one of the two settings in the DIAG menu that is retained after you exit the
menu. To access this feature press the following keys. Also note that the ozone generator
does not turn on if the ozone flow conditions are out of specification (e.g., if there is no flow
through the system or the pump is broken).
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR EXIT
DIAG
SIGNAL I / O
PREV NEXT JUMP
ENTR
EXIT
DIAG OZONE OZONE GEN OVERRIDE
OFF
ENTR EXIT
Press NEXT until…
DIAG OZONE GEN OVERRIDE
PREV NEXT
ENTR EXIT
Toggle this key to turn the O
3
generator ON/OFF.