To activate the electrical test press the following keys.

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS DIAG

EXIT

DIAG

SIGNAL I / O


PREV NEXT JUMP

ENTR

EXIT

DIAG ELEC RANGE = 500.0 PPB NOX=X.X

EXIT

Press NEXT until…

DIAG

ELECTRICAL TEST


PREV NEXT

ENTR EXIT

Press TST until…

DIAG ELEC PMT = 1732 MV NOX=X.X


EXIT

While the electrical test is

activated, PMT should equal:

2000 mV ± 1000 mV

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR EXIT

6.7.7. Ozone Generator Override

This feature allows the user to manually turn the ozone generator off and on. This can be
done before disconnecting the generator, to prevent ozone from leaking out, or after a

system restart if the user does not want to wait for 30 minutes during warm-up time. Note

that this is one of the two settings in the DIAG menu that is retained after you exit the

menu. To access this feature press the following keys. Also note that the ozone generator

does not turn on if the ozone flow conditions are out of specification (e.g., if there is no flow

through the system or the pump is broken).

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS DIAG

EXIT

SETUP X.X

ENTER DIAG PASS: 818


8 1 8

ENTR EXIT

DIAG

SIGNAL I / O


PREV NEXT
JUMP

ENTR

EXIT

DIAG OZONE OZONE GEN OVERRIDE

OFF

ENTR EXIT

Press NEXT until…

DIAG OZONE GEN OVERRIDE

PREV NEXT

ENTR EXIT

Toggle this key to turn the O

3

generator ON/OFF.