Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 107
Model 9110E Instruction Manual
Operating Instructions
M9110E Rev B
93
Table 0-22: M9110E Default iDAS Configuration, Revision C.7.
List of Channels
List of Parameters
NAME – MODE/EVENT - PRECISION - NUM SAMPLES
NOXCNC1 - AVG - 4 - ON
NOCNC1 - AVG - 4 - OFF
N2CNC1 - AVG - 4 - OFF
Name: CONC
Event: ATIMER
Sample Period: 000:00:01
Report Period: 000:01:00
Number of Records: 800
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: ON
STABIL - AVG - 4 - OM
NXZSC1 - SLPCHG - 4 - OFF
NOXSLP1 - SLPCHG - 4 - OFF
NOXOFFS1 - SLPCHG - 4 - OFF
NOZSC1 - SLPCHG - 4 - OFF
NOSLP1 - SLPCHG - 4 - OFF
NOOFFS1 - SLPCHG - 4 - OFF
N2ZSC1 - SLPCHG - 4 - OFF
CNVEF1 - SLPCHG - 4 - OFF
Name: CALDAT
Event: SLPCHG
Number of Records: 200
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
STABIL - SLPCHG - 4 - OFF
NXZSC1 - EXITMP - 4 - OFF
NOZSC1 - EXITMP - 4 - OFF
N2ZSC1 - EXITMP - 4 - OFF
Name: CALCHECK
Event: EXITMP
Number of Records: 200
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
STABIL - EXITMP - 4 - OFF
SMPFLW - AVG – 2 - OFF
O3FLOW – AVG – 2 - OFF
RCPRESS – AVG – 2 - OFF
SMPPRES – AVG – 2 - OFF
RCTEMP - AVG – 2 - OFF
PMTTMP - AVG – 2 - OFF
CNVTMP - AVG – 2 - OFF
BOXTMP - AVG – 2 - OFF
HVPS - AVG – 2 - OFF
Name: DIAG
Event: ATIMER
Sample Period: 000:00:30
Report Period: 001:00:00
Number of Records: 1100
RS-232 report: OFF
Channel enabled: ON
DAS HOLDOFF: OFF
AZERO - AVG – 2 - OFF
NOXCNC1 - AVG - 4 - OFF
NOCNC1 – AVG - 4 - OFF
N2CNC1 - AVG – 4 - OFF
STABIL - AVG – 4 - OFF
SMPFLW - AVG – 2 - OFF
O3FLOW - AVG – 2 - OFF
RCPRESS - AVG – 2 - OFF
SMPPRES - AVG – 2 - OFF
RCTEMP - AVG – 2 - OFF
PMTTMP - AVG – 2 - OFF
CNVTMP - AVG – 2 - OFF
BOXTMP - AVG – 2 - OFF
HVPS - AVG – 1 - OFF
AZERO - AVG – 2 - OFF
REFGND - AVG – 1 - OFF
Name: HIRES
Event: ATIMER
Sample Period: 000:00:01
Report Period: 000:00:01
Number of Records: 1500
RS-232 report: OFF
Channel enabled: OFF
DAS HOLDOFF: OFF
REF4096 - AVG - 1 - OFF
DIAG: Daily averages of temperature zones, flow and pressure data as well as some other
diagnostic parameters (HVPS, AZERO). These data are intended for predictive diagnostics
and maintenance described in Chapter 9. The last 1100 daily averages are stored to cover
more than four years of analyzer performance, a major portion of the analyzer’s lifetime. If
these parameters are required for data analysis, this channel can easily be duplicated (e.g.,