APPENDIX A

Model 9110E Instruction Manual

242

044100120 Rev B

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

TEST_CHAN_ID —

NONE NONE,

PMT DE-
TECTOR,

OZONE

FLOW,
SAMPLE

FLOW,

SAMPLE
PRESSUR

E,

RCELL

PRESSUR
E,

RCELL

TEMP,
MANIFOL

D TEMP,

IZS
TEMP,

CONV

TEMP,

PMT
TEMP,

CHASSIS

TEMP,
HVPS

VOLTAGE

Diagnostic analog output ID.

Enclose value in double quotes (")
when setting from the RS-232

interface.

COMMAND_PROMPT —

“Cmd>

Any
character

in the

allowed

character
set. Up

to 100

character
s long.

RS-232 interface command
prompt. Displayed only if enabled

with RS232_MODE variable.

Enclose value in double quotes (")

when setting from the RS-232
interface.

REMOTE_CAL_MODE

3

LOW

LOW,

HIGH

Range to calibrate during remote

calibration. Enclose value in double
quotes (") when setting from the

RS-232 interface.

PASS_ENABLE

OFF

ON, OFF

ON enables passwords; OFF
disables them.

STABIL_GAS — NOX NO,

NO2,
NOX

Selects gas for stability

measurement. Enclose value in
double quotes (") when setting

from the RS-232 interface.

STABIL_FREQ Seconds

10

1-300

Stability measurement sampling
frequency.

STABIL_SAMPLES

Samples

25

2-40

Number of samples in concentra-
tion stability reading.

HVPS_SET Volts

650

Warnings:
400-900

0-2000

High voltage power supply warning

limits. Set point is not used.