Model 9110E Instruction Manual

Operating Instructions

M9110E Rev B

91

Property Description

Default

Setting

Range

channel values to the RS-232 ports.

COMPACT
REPORT

Enables a compact version of the RS-232
reporting with up to five parameters sharing

the same output line.

OFF OFF

or

ON

CHANNEL
ENABLED

Enables or disables the channel. Provides a
convenient means to temporarily disable a

data channel.

ON OFF

or

ON

CAL HOLD OFF

Disables sampling of data parameters while

instrument is in calibration mode.
Note that - when enabled here - there is

also a length of the DAS HOLD OFF after
calibration mode, which is set in the VARS

menu. (Section 6.5.)

OFF OFF

or

ON

6.10.1.2. iDAS Parameters

Data parameters are types of data that may be measured and stored by the iDAS. For each

TAI analyzer model, the list of available data parameters is different, fully defined and not

customizable. Appendix A-5 lists firmware specific data parameters for the M9110E. The

most common parameters are concentrations of measured gases (NO

X

, NO and NO

2

),

temperatures of heated zones (NO

2

converter, reaction cell, box temperature…), pressures

and flows of the pneumatic subsystem and other diagnostic measurements as well as

calibration data (slope and offset) for each gas.

Most data parameters have associated measurement units, such as mV, ppb, cm³/min, etc.,

although some parameters have no units. The only units that can be changed are those of

the concentration readings according to the SETUP-RANGE settings. Note that the iDAS

does not keep track of the unit of each concentration value and iDAS data files may contain

concentrations in multiple units if the unit was changed during data acquisition.

Each data parameter has user-configurable functions that define how the data are recorded

(Table 0-21).

Table 0-21: iDAS Data Parameter Functions

Function Effect

PARAMETER

Instrument-specific parameter name.

SAMPLE MODE

INST: Records instantaneous reading.
AVG: Records average reading during reporting interval.
MIN: Records minimum (instantaneous) reading during reporting interval.
MAX: Records maximum (instantaneous) reading during reporting interval.

PRECISION

Decimal precision of parameter value(0-4).

STORE NUM. SAMPLES OFF: stores only the average (default).

ON: stores the average and the number of samples in each average for a

parameter. This property is only useful when the AVG sample mode is used.

Note that the number of samples is the same for all parameters in one
channel and needs to be specified only for one of the parameters.

Users can specify up to 50 parameters per data channel (the M9110E has about 30 para-
meters). The number of parameters and channels is also limited by analyzer memory.