Generator replacement, Sample and ozone dryer replacement – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 236

background image

Troubleshooting & Repair

Model 9110E Instruction Manual

222

M9110E Rev B

• When looking at the electronic circuits from the back of the analyzer, locate the

EEPROM chip in the left-most socket of the CPU board. The chip is almost square

with one corner cut off, the socket is shaped accordingly and the chip is recessed

into the socket. Remove the old chip by using a special tool or gently pry the chip

out using a very fine screwdriver. Make sure not to bend or destroy any of the

contacts of the socket. When upgrading the CPU with a second chip, no removal is

necessary as the second socket should be empty.

• Reinstall the new or additional EEPROM chip, making sure the cut-off edge matches

that of the socket. Press the chip symmetrically and straight all the way in.

• Close the rear panel and cover and turn on power to the machine.
• If a front panel message Flash Format INVALID appears on start-up, the EEPROM

was not properly formatted. Contact the factory for a proper replacement.

11.6.3. O

3

Generator Replacement

The ozone generator is a black, brick-shaped device with printed circuit board attached to

its rear and two tubes extending out the right side in the front of the analyzer. To replace
the ozone generator:

• Turn off the analyzer power, remove the power cord and the analyzer cover.
• Disconnect the 1/8” black tube from the white ozone scrubber cartridge and the ¼”

clear tube from the plastic extension tube at the brass fitting nearest to the ozone

generator. Unplug the electrical connection on the rear side of the brick.

• Unscrew the two mounting screws that attach the ozone generator to the chassis

and take out the entire assembly.

• If you received a complete replacement generator with circuit board and mounting

bracket attached, simply reverse the above steps to replace the current generator.
Make sure to carry out a leak check and a recalibration after the analyzer warmed up

for about 30 minutes.

11.6.4. Sample and Ozone Dryer Replacement

The M9110E standard configuration is equipped with a dryer for the ozone supply air. An

optional dryer is available for the sample stream and a combined dryer for both gas
streams can also be purchased. To change one or all of these options:

• Turn off power to the analyzer and pump, remove the power cord and the analyzer

cover. Locate the dryers in the center of the instrument, between sensor and NO

2

converter. They are mounted to a bracket, which can be taken out when unscrewing

the two mounting screws (if necessary).

• Disconnect all tubing that extends out of the dryer assembly, those are usually the

purge tube connecting to the vacuum manifold, the tube from the exit to the ozone
flow meter (ozone dryer) or to the NO/NO

x

valve (sample dryer) or two tubes to the

ozone flow meter and the NO/NO

X

valve (combo-dryer). Take extra care not to twist

any of the white plastic fittings on the dryer, which connect the inner drying tube to

the outer purge tube. See Section 9.3.2 and Figure 9-2.

• Note the orientation of the dryer on the bracket. Cut the tie wraps that hold the

dryer to the mounting bracket and take out the old dryer. If necessary, unscrew the

two mounting screws on the bracket and take out the entire assembly.