Operating Instructions

Model 9110E Instruction Manual

72

M9110E Rev B

Table 0-12: Current Loop Output Calibration with Resistor

Full scale

Voltage for 2-20 mA

(measured across resistor)

Voltage for 4-20 mA

(measured across resistor)

0% 0.5

V

1.0

V

100% 5.0

V

5.0

V

6.7.3.6. AIN Calibration

This is the sub-menu to conduct the analog input calibration. This calibration should only be
necessary after major repair such as a replacement of CPU, motherboard or power supplies.



Exit at any time to

return to the main

DIAG menu

DIAG

ANALOG I / O CONFIGURATION


PREV

NEXT

ENTR

EXIT

DIAG AIO

AIN A/C FREQUENCY: 60 HZ


SET>
EDIT

EXIT

DIAG AIO

AIN CALIBRATED: NO


< SET SET> CAL

EXIT

DIAG AIO

AIN CALIBRATED: YES


< SET SET> CAL

EXIT

Instrument

calibrates

automatically

Exit to return to the

ANALOG I/O

CONFIGURATION

MENU

STARTING FROM ANALOG I / O CONFIGURATION MENU

DIAG AIO

CALIBRATING A/D ZERO


CALIBRATING A/D SPAN

6.7.4. Test Channel Output

When activated, output channel A4 can be used to report one of the test functions viewable

from the SAMPLE mode display.