beautypg.com

Checking for light leaks – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 172

background image

Instrument Maintenance

Model 9110E Instruction Manual

158

M9110E Rev B

Figure 9-7: Critical Flow Orifice Assembly

9.3.11. Checking for Light Leaks

When re-assembled or operated improperly, the M9110E can develop small leaks around

the PMT, which let stray light from the analyzer surrounding into the PMT housing. To find
such light leaks, follow the below procedures. CAUTION: this procedure can only be carried

out with the analyzer running and its cover removed. This procedure should only be carried
out by qualified personnel.

• Scroll the TEST functions to PMT.
• Supply zero gas to the analyzer.
• With the instrument still running, carefully remove the analyzer cover. Take extra

care not to touch any of the inside wiring with the metal cover or your body. Do not
drop screws or tools into a running analyzer!

• Shine a powerful flashlight or portable incandescent light at the inlet and outlet

fitting and at all of the joints of the reaction cell as well as around the PMT housing.

The PMT value should not respond to the light, the PMT signal should remain steady

within its usually noise.

• If there is a PMT response to the external light, symmetrically tighten the reaction

cell mounting screws or replace the 1/4” vacuum tubing with new, black PTFE tubing
(this tubing will fade with time and become transparent). Often, light leaks are also

caused by O-rings being left out of the assembly.

• Carefully replace the analyzer cover. If tubing was changed, carry out a leak check

(Section 11.5.2).