Calibration Procedures
Model 9110E Instruction Manual
122
M9110E Rev B
Wait until STABIL
value falls below
0.5 ppb. This may
take several
minutes.
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X
< TST TST > ZERO CONC
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X
< TST TST > ENTR CONC
EXIT
If the ZERO button does
not appear, see Chapter
11 for troubleshooting.
Analyzer enters ZERO CAL
mode
EXIT terminates
calibration and returns to
SAMPLE mode.
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X
< TST TST > ZERO CONC
EXIT
ZERO CAL M NOX STB=XXX.X PPB NOX=X.X X
< TST TST > CAL CALZ CALS
SETUP
SAMPLE RANGE = 500.0 PPB NOX= X.X
< TST TST > CAL CALZ CALS
SETUP
SAMPLE NOX STB=XXX.X PPB NO = X.X
< TST TST > CAL CALZ CALS
SETUP
Set the Display to show the
NOX STB test function.
Pressing ENTR changes the
calibration of the instrument.
Turn ON the CAL_ON_NO
2
feature:
SAMPLE RANGE = 500.0 PPB NOX=X.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
EXIT
SETUP X.X
0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
ENTER VARS PASS: 818
8 1 8
ENTR EXIT
EXIT ignores the new setting.
ENTR accepts the new setting.
SETUP X.X
8) CAL_ON_NO
2
= OFF
PREV NEXT JUMP
EDIT PRNT EXIT
SETUP X.X
8) CAL_ON_NO
2
= ON
ON
ENTR EXIT
Toggle this key to change setting
Press NEXT until …