Model 9110E Instruction Manual

Operating Instructions

M9110E Rev B

49

Chapter 7 details all calibration operations, Chapter 1 describes the zero/span valve and

IZS options.

6.3. Calibration Mode

Owing to its importance, calibration is described separately in Chapter 7, EPA protocol

calibration in Chapter 8.

6.4. Setup Mode

The SETUP mode contains a variety of choices that are used to configure the analyzer’s
hardware and software features and to read or configure the internal data acquisition

system (iDAS). For a visual representation of the software menu trees, refer to
Appendix A-1.

NOTE

Any changes made to a variable during one of the following procedures is not

acknowledged by the instrument until the ENTR Key is pressed

If the EXIT key is pressed before the ENTR key, the analyzer will beep, alerting the user

that the newly entered value has not been accepted.

Some of the functions of the SETUP mode are described in separate sections following this

section. In particular, the RNGE function can be found in Section 6.5, the ACAL function in

Section 7.8.

6.4.1. Password (PASS)

The M9110E provides password protection of the calibration and setup functions to prevent

unauthorized adjustments. When the passwords have been enabled in the PASS menu

item, the system will prompt the user for a password anytime a password-protected
function is requested.

There are three levels of password protection, which correspond to operator, maintenance,
and configuration functions. Each level allows access to all of the functions in the previous

level.

Table 0-4: Password Levels

Password Level

Menu Access Allowed

No password

Operator

TEST, MSG, CLR

101

Maintenance

CAL, CALZ, CALS

818

Configuration

SETUP, VARS, DIAG