Model 9110E Instruction Manual

Operating Instructions

M9110E Rev B

103

To enable automatic COM port reporting, press SETUP-DAS-ENTR and then:


Edit Data Channel Menu

SETUP X.X 0) CONC: ATIMER, 2, 900

PREV NEXT INS DEL EDIT PRNT EXIT

FROM THE DAS PASSWORD MENU

Exits to the main

Data Acquisition

menu

SETUP X.X NAME:CONC

SET> EDIT PRINT

EXIT

SETUP X.X RS-232 REPORT: OFF

EDIT PRINT

EXIT

ENTR accepts the new

setting and returns to the

previous menu.

EXIT ignores the new setting

and returns to the previous

menu.

SETUP X.X RS-232 REPORT: OFF

OFF

ENTR EXIT

Toggle key to turn

reporting ON or OFF

Press SET> key until…

6.10.2.9. Compact Report

When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of

reporting each parameter in one channel on a separate line, up to five parameters are

reported in one line, instead. For example, channel DIAG would report its record in two lines

(10 parameters) instead of 10 lines. Individual lines carry the same time stamp and are
labeled in sequence.

6.10.2.10. Starting Date

This option allows to specify a starting date for any given channel in case the user wants to

start data acquisition only after a certain time and date. If the Starting Date is in the past,
the iDAS ignores this setting.

6.10.2.11. Disabling/Enabling Data Channels

Data channels can be temporarily disabled, which can reduce the read/write wear on the
disk-on-chip. The HIRES channel of the M9110E, for example, is disabled by default.