Troubleshooting & Repair

Model 9110E Instruction Manual

196

M9110E Rev B

voltage of an input signal or to control the state of an output voltage or control signal. The

specific parameter will vary depending on the situation.

SAMPLE

RANGE = 500.0 PPB

NOX=XXX.X

SETUP

CFG DAS RNGE PASS CLK MORE

SETUP

COMM VARS DIAG HALT

Exit returns to

DIAG display & all

values return to software

DIAG

SIGNAL I/O


PREV NEXT

ENTR EXIT

If parameter is an

input signal

DIAG I/O

0 ) EXT_ZERO_CAL=ON


PREV NEXT JUMP

PRNT

DIAG I/O 19 ) REACTION
CELL_HEATER=ON

SETUP ENTER DIAG PASS: 818

8 1 8

ENTR

DIAG I/O 29) PMT_TEMP=378.3 MV

PREV NEXT JUMP

PRNT

DIAG I/O 19 ) REACTION
CELL_HEATER=OFF

If parameter is an

output signal or

Toggles

parameter

ON/Off

Press ENTR to

enter the

Press ENTR again

to enter the

SIGNAL I/O menu

Press PREV or

NEXT to scroll

between available

I/O parameters

Figure 11-2: Switching Signal I/O Functions

11.1.4. Status LED’s

Several color-coded, light-emitting diodes (LED) are located inside the instrument to

determine if the analyzer’s CPU, I

2

C communications bus and the relay board are

functioning properly.

11.1.4.1. Motherboard Status Indicator (Watchdog)

A red LED labeled DS5 in the upper portion of the motherboard (Figure 11-3), just to the

right of the CPU board, flashes when the CPU is running the main program. After power-up,

DS5 should flash on and off about once per second. If characters are visible on the front
panel display but DS5 does not flash then the program files have become corrupted.

Contact customer service because it may be possible to recover operation of the analyzer. If

30 - 60 seconds after a restart neither DS5 is flashing nor any characters are visible on the

front panel display, the firmware may be corrupted or the CPU may be defective. If DS5 is

permanently off or permanently on, the CPU board is likely locked up and the analyzer

should not respond (either with locked-up or dark front panel).