11.6.6. Replacing the PMT, HVPS or TEC

The photo multiplier tube (PMT) should last for the lifetime of the analyzer. However, in

some cases, the high voltage power supply (HVPS) or the thermo-electric cooler (TEC) may

fail. In case of PMT, HVPS or TEC failure, the sensor assembly needs to be opened in order
to change one of these components. Refer to Figure 11-7 for the structure of the 9110E

sensor assembly and follow the steps below for replacement of one of its components. We
recommend to ensure that the PMT, HVPS or TEC modules are, indeed, faulty to prevent

unnecessary opening of the sensor.

• Power down the analyzer, disconnect the power cord, remove the cover and discon-

nect all pneumatic and electrical connections from the sensor assembly.

• Whereas it is possible for a skilled technician to change the PMT or HVPS through the

front panel with the sensor assembly mounted to the analyzer, we recommend to

remove the entire assembly and carry this procedure out on a clean, anti-static table
with the user wearing an anti-static wrist strap to prevent static discharge damage

to the assembly or its circuits.

• If the TEC is to be replaced, remove the reaction cell assembly at this point by

unscrewing two holding screws. This is necessary only if the PMT cold block is to be

removed. This step is not necessary if the HVPS or the PMT only are exchanged.

Figure 11-7: M9110E Sensor Assembly

• Remove the two connectors on the PMT housing end plate facing towards the front

panel. Remove the end plate itself (4 screws with plastic washers). Remove the

dryer packages inside the PMT housing. Along with the plate, slide out the OPTIC

TEST LED and the thermistor that measures the PMT temperature. Both may be