Status reporting – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 122

background image

Operating Instructions

Model 9110E Instruction Manual

108

M9110E Rev B

executed. Press ? or refer to Appendix A-6 for a list of available

command designators.

is a carriage return. All commands must be terminated by a carriage

return (usually achieved by pressing the ENTER key on a computer).

Table 0-24: Command Types

Command Command

Type

C Calibration
D Diagnostic

L Logon
T Test

measurement

V Variable

W Warning

6.11.1.4. Data Types

Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean
expressions and text strings.

• Integer data are used to indicate integral quantities such as a number of records, a

filter length, etc. They consist of an optional plus or minus sign, followed by one or

more digits. For example, +1, -12, 123 are all valid integers.

• Hexadecimal integer data are used for the same purposes as integers. They consist

of the two characters “0x,” followed by one or more hexadecimal digits (0-9, A-F, a-

f), which is the ‘C’ programming language convention. No plus or minus sign is
permitted. For example, 0x1, 0x12, 0x1234abcd are all valid hexadecimal integers.

• Floating-point numbers are used to specify continuously variable values such as

temperature set points, time intervals, warning limits, voltages, etc. They consist of

an optional plus or minus sign, followed by zero or more digits, an optional decimal
point, and zero or more digits. (At least one digit must appear before or after the

decimal point.) Scientific notation is not permitted. For example, +1.0, 1234.5678, -

0.1

, 1 are all valid floating-point numbers.

• Boolean expressions are used to specify the value of variables or I/O signals that

may assume only two values. They are denoted by the keywords ON and OFF.

• Text strings are used to represent data that cannot be easily represented by other

data types, such as data channel names, which may contain letters and numbers.
They consist of a quotation mark, followed by one or more printable characters,

including spaces, letters, numbers, and symbols, and a final quotation mark. For

example, “a”, “1”, “123abc”, and “()[]<>” are all valid text strings. It is not possible

to include a quotation mark character within a text string.

• Some commands allow you to access variables, messages, and other items, such as

iDAS data channels, by name. When using these commands, you must type the

entire name of the item; you cannot abbreviate any names.

6.11.2. Status Reporting

Reporting of status messages as an audit trail is one of the three principal uses for the

RS-232 interface (the other two being the command line interface for controlling the