Hessen mode configuration – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 102

background image

Operating Instructions

Model 9110E Instruction Manual

88

M9110E Rev B

Step three in the configuration process is to restart the pass-through mode for normal

operation.


SETUP X.X

PRIMARY SETUP MENU


CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS DIAG

EXIT

SAMPLE RANGE = 500.0 PPB O3 =XXX.X

< TST TST > CAL

SETUP

ETHERNET STARTED

SETUP X.X COMMUNICATIONS MENU

ID INET COM1 COM2

EXIT

INTERNET CONFIGURATION


GTWY IP SNET PORT STRT STOP EXIT

STARTING ETHERNET

If the message ETHERNET NOT STARTED

appears here, the Ethernet card either was already

in configuration mode or might not be responding.

6.9.10. Hessen Mode Configuration

The Hessen protocol is used in applications, which adhere to operation protocols originally
defined by the German state of Hessen and also applies to the Bavarian protocol.

A Hessen protocol network typically consists of a host computer that monitors the status

and controls one or more remote instruments. The remote instruments are regarded as

slaves of the host computer, and only respond to commands from the host computer;

remote instruments never initiate Hessen protocol messages.

The Hessen protocol is a multidrop application, in which several instruments are connected

to a common communications channel (e.g., using hardware Option 62). However, the
implementation details of this multidrop bus are handled outside of the instrument. The

instruments are unaware that they are connected to a multidrop bus and respond only to

commands containing their own unique ID number.

The Hessen protocol is designed to accomplish two things:

• to obtain the status of remote instruments, including the concentrations of all the

gases measured and

• to place remote instruments into zero or span calibration or measure mode.

The TAI implementation supports both of these principal features.

From the user’s point of view, the RS-232 hardware on TAI instruments supporting the

Hessen protocol is not different from that of standard instruments. The RS-232

communication parameters, however, are different from standard configurations. For more

information on configuring the Hessen protocol, please download the manual addendum at

http://www.teledyne-api.com/manuals/.