Operating Instructions

Model 9110E Instruction Manual

74

M9110E Rev B

6.7.5. Optic Test

The optic test function tests the response of the PMT sensor by turning on an LED located in

the cooling block of the PMT (Fig. 10-15). The analyzer uses the light emitted from the LED

to test its photo-electronic subsystem, including the PMT and the current to voltage
converter on the pre-amplifier board. To make sure that the analyzer measures only the

light coming from the LED, the analyzer should be supplied with zero air. The optic test
should produce a PMT signal of about 2000±1000 mV. To activate the electrical test press

the following key sequence.

SAMPLE RANGE = 500.0 PPB NOX=X.X

< TST TST > CAL

SETUP

SETUP X.X

PRIMARY SETUP MENU


CFG DAS RNGE PASS CLK MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU


COMM VARS DIAG

EXIT

DIAG

SIGNAL I / O


PREV NEXT
JUMP

ENTR

EXIT

DIAG OPTIC RANGE = 500.0 PPB NOX=X.X

EXIT

SETUP X.X ENTER DIAG PASS: 818

8 1 8

ENTR EXIT

Press NEXT until…

DIAG

OPTIC TEST


PREV NEXT

ENTR EXIT

Press TST until…

DIAG ELEC PMT = 2751 MV NOX=X.X


EXIT

While the optic test is

activated, PMT should be

2000 mV ± 1000 mV

NOTE

This is a coarse test for functionality and not an accurate calibration tool. The

resulting PMT signal can vary significantly over time and also changes with low-

level calibration.

6.7.6. Electrical Test

The electrical test function creates a current, which substitutes the PMT signal, and feeds it

into the preamplifier board. This signal is generated by circuitry on the pre-amplifier board

itself and tests the filtering and amplification functions of that assembly along with the A/D

converter on the motherboard. It does not test the PMT itself. The electrical test should
produce a PMT signal of about 2000 ±1000 mV.