Operating Instructions

M9110E Rev B

59

6.6. Setup - Internal Variables (VARS)

The M9110E has several-user adjustable software variables, which define certain

operational parameters. Usually, these variables are automatically set by the instrument’s

firmware, but can be manually re-defined using the VARS menu. Table 0-5 lists all variables

that are available within the 818 password protected level. See Appendix A2 for a detailed

listing of all of the M9110E variables that are accessible through the remote interface.

Table 0-5: Variable Names (VARS) Revision C.4

Variable Description

Allowed

Values

DAS_HOLD_OFF

Duration of no data storage in the iDAS. This is the time

when the analyzer returns from one of its calibration
modes to the SAMPLE mode. The DAS_HOLD_OFF can be

disabled in each iDAS channel.

Can be between 0.5

and 20 minutes
Default=15 min.

TPC_ENABLE

Enables or disables the temperature and pressure
compensation (TPC) feature (Section 10.4.3).

ON/OFF
Default=ON

DYN_ZERO

Dynamic zero automatically adjusts offset and slope of

the NO and NO

X

response when performing a zero point

calibration during an AutoCal (Chapter 7).

ON/OFF

Default=OFF

DYN_SPAN

Dynamic span automatically adjusts the offsets and
slopes of the NO and NOx response when performing a

zero point calibration during an AutoCal (Chapter 7).
Note that the DYN_ZERO and DYN_SPAN features are not

allowed for applications requiring EPA equivalency.

ON/OFF
Default=OFF

SFLOW_SET

Adjusts the displayed value of the sample flow rate to the
actual flow rate, which was measured by the user.

Changing this value does not alter the actual sample gas

flow rate!

0-1000 cm³/min
Default=

500 cm³/min

OFLOW_SET

Adjusts the displayed value of the ozone flow rate to its

actual flow rate, which was measured by the user.

Changing this value does not alter the actual ozone gas
flow rate!

10-1000 cm³/min

Default=

80 cm³/min

IZS_SET

Sets the IZS oven temperature. Increasing or decreasing
this temperature will increase or decrease the NO

2

permeation rate of the IZS source (Section 5.6.3).

30

°C - 70°C

Default= 50°C

CONC_PRECISION Allows to set the number of decimal points of the

concentration and stability parameters displayed on the

front panel.

AUTO, 1, 2, 3, 4
Default=AUTO

CLOCK_ADJ

Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock

is too fast.

-60 to +60 s/day
Default=0

CAL_ON_NO

2

Allows to turn ON and OFF the ability to span the
analyzer with NO

2

, in which case the instrument acts as if

NO and NO

X

are spanned, even though it is supplied with

NO

2

. The NO

2

concentration is then zero by default.

This feature is not permitted for EPA protocol

applications.

ON or OFF
Default=OFF