Warm-up – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 35
Model 9110E Instruction Manual
Getting Started
M9110E Rev B
21
Firmware
fully booted
The revision level of the firmware
installed in your analyzer is briefly
displayed
SOFTWARE REVISION C.4
BOOT PROGRESS [XXXXXXXX 80% _ _]
The instrument is loading the
analyzer firmware.
9110E NOX ANALYZER
BOOT PROGRESS [XXXXX 50%_ _ _ _ _]
The instrument is loading
configuration and calibration data
from the flash chip
STARTING INSTRUMENT W/FLASH : 1
System waits 3 seconds then
automatically begins its initialization
routine. No action required.
System is checking the format of
the instrument’s flash memory chip.
SELECT DESIRED FUNCTION : 3
START
.
CHECKING FLASH STATUS
: 2
STARTING INSTRUMENT CODE : 1
SAMPLE SYSTEM RESET NOX=X.X
TEST CAL CLR SETUP
If at this point,
**FLASH FORMAT INVALID**
appears, contact T–API customer
service
Press CLR to clear initial warning messages.
(see Section 3.2.3)
Figure 3-10: Front Panel Display During Startup Sequence
3.5.2. Warm-Up
The M9110E requires about 30 minutes warm-up time before reliable NO
X
, NO and NO
2
measurements can be taken. During that time, various portions of the instrument’s front
panel will behave as follows. See Figure 3-4 for locations.
Table 3-5: Front Panel Display During System Warm-Up
Name Color Behavior
Significance
Concentra-
tion Field
N/A
Switches between
NO
X
, NO and NO
2
This is normal operation.
Mode Field
N/A
Displays blinking
“SAMPLE”
Instrument is in sample mode but is still in the
process of warming up (hold-off period is
active).