beautypg.com

Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 258

background image

APPENDIX A

Model 9110E Instruction Manual

244

044100120 Rev B

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

CLOCK_FORMAT —

“TIME=%H:

%M:%S”

Any

character
in the

allowed

character
set. Up

to 100

charac-
ters long.

Time-of-day clock format flags.

Enclose value in double quotes (")
when setting from the RS-232

interface.

“%a” = Abbreviated weekday
name.

“%b” = Abbreviated month name.

“%d” = Day of month as decimal
number (01 – 31).

“%H” = Hour in 24-hour format

(00 – 23).

“%I” = Hour in 12-hour format (01
– 12).

“%j” = Day of year as decimal

number (001 – 366).
“%m” = Month as decimal number

(01 – 12).

“%M” = Minute as decimal number
(00 – 59).

“%p” = A.M./P.M. indicator for 12-

hour clock.

“%S” = Second as decimal number
(00 – 59).

“%w” = Weekday as decimal

number (0 – 6; Sunday is 0).
“%y” = Year without century, as

decimal number (00 – 99).

“%Y” = Year with century, as
decimal number.

“%%” = Percent sign.

FACTORY_OPT

0

0-65535

Factory option flags. Add values to

combine flags.
1 = enable dilution factor
2 = display units in concentration
field
4 = zero/span valves installed
8 = low span valve installed
16 = IZS and zero/span valves
installed
32 = enable software-controlled

maintenance mode
64 = display temperature in

converter warning message
128 = enable switch-controlled
maintenance mode
256 = enable simultaneous display
of all gas concentrations
512 = enable manifold tempera-

ture control

1

Optional

Equipment