EPA Protocol Calibration

Model 9110E Instruction Manual

132

M9110E Rev B

Table 8-1: Activity Matrix for EPA Calibration Equipment and Supplies

Equipment/

Supplies

Acceptance Limits

Frequency And Method

Of Measurement

Action If Requirements

Are Not Met

Recorder

Compatible with output signal
of analyzer; min chart width of

150 mm (6 in) is recommended

Check upon receipt

Return equipment to
supplier

Sample line
and manifold

Constructed of PTFE, glass or
stainless steel

Check upon receipt

Return equipment to
supplier

Calibration

equipment

Meets guidelines and Section

2.3.2 of Q. A. Handbook

See Section 2.3.9 of Q.

A. Handbook

Return equipment/

supplies to supplier or
take corrective action

Working

standard NO
cylinder gas

Traceable to NIST-SRM. Meets

limits in traceability protocol for
accuracy and stability.

Section 2.0.7 of Q. A. Handbook

Analyzed against NIST-

SRM; see protocol in
Section 2.0.7, Q.A.

Handbook

Obtain new working

standard and check for
traceability

Recording

forms

Develop standard forms

N/A

Revise forms as

appropriate

Audit

equipment

Cannot be the same as used for

calibration

System must be

checked out against
known standards

Locate problem and

correct or return to
supplier

Table 8-2: Activity Matrix for Calibration Procedure

Equipment/

Supplies

Acceptance Limits

Frequency And Method

Of Measurement

Action If Requirements

Are Not Met

Calibration

gases

Sec. 2.0.7, Q.A. Handbook

Assayed against an

NIST-SRM quarterly,
Sec. 2.0.7

(Q.A. Handbook)

Working gas standard is

unstable, and/or
measurement method is

out of control; take

corrective action such as
obtaining new calibration

gas

Dilution gas

Zero air, free of contaminants;

TAD

2

and Sec. 2.0.7, Subsec.

7.1, Q.A. Handbook

See TAD

2

Return to supplier or take

appropriate action with
generation system

Multi-point

calibration
(GPT)

t

R

< 2 minutes P

R

> 2.75 ppm/min
Use calibration procedure in Q.A

Handbook; also CFR

1

Converter efficiency > 96%

Method: Section 7.2 of

this manual)
CFR Title 40

1

,

Appendix F

Adjust flow conditions

and/or reaction cell
volume to meet

suggested limits
Repeat the calibration
Replace or service the

converter