Calibration with independent or auto ranges, Calibration in auto range mode – Teledyne 9110E - Nitrogen Oxides Analyzer User Manual

Page 138

background image

Calibration Procedures

Model 9110E Instruction Manual

124

M9110E Rev B

• Perform the zero/span check. Zero and span checks using the zero/span valve or

IZS option are similar to that described in Section 7.3, except:

On units with an IZS option installed, zero air is supplied to the analyzer through the zero

gas inlet and the same gas stream is used to create span gas internally using the permea-

tion device.

On units with a zero/span valve option installed, zero air and span gas are supplied to the

analyzer through the zero gas and span gas inlets from two different sources.

The zero and calibration operations are initiated directly and independently with dedicated

keys CALZ and CALS.

Note that there is no option to use a dedicated span gas port if the IZS option is installed.

Span gas from a gas tank would need to be supplied through the Sample port or through

one of the auxiliary ports available on the rear panel (with modified internal plumbing),

although the user could modify the plumbing to have the span gas go through the IZS oven

(without the permeation tube installed).

Wait until NOX

STB falls below

0.5 ppb. This may

take several

minutes.

The value of NOX

STB may jump

significantly. Wait

until NOX STB falls
below 0.5 ppb. This

may take several

minutes.

SPAN CAL M NOX STB=XXX.X PPB NOX=X.XX

< TST TST > ZERO SPAN CONC

EXIT

SAMPLE NOX STB=XXX.X PPB NOX=X.X

< TST TST > CAL CALZ CALS

SETUP

ZERO CAL M NOX STB=XXX.X PPB NOX=X.X

< TST TST > ZERO CONC

EXIT

SAMPLE NOX STB=XXX.X PPB NOX=X.X

< TST TST > CAL CALZ CALS

SETUP

EXIT returns to the main

SAMPLE display

ACTION:

Record the

NO

X

, NO

2

& NO

readings presented in
the upper right corner

of the display.

ACTION:

Record the

NO

X

, NO

2

& NO

readings presented in
the upper right corner

of the display.

SAMPLE RANGE = 500.0 PPB NOX=X.XX

< TST TST > CAL CALZ CALS

SETUP

Scroll to the NOX

STB test function.

7.7. Calibration With Independent or AUTO Ranges

Additional considerations apply when calibrating a Model 9110E with reporting ranges set to

AUTO (automatic) or IND (independent) mode (Sections 6.5.4 and 6.5.5).

7.7.1. Calibration in AUTO Range Mode

If the analyzer is operated in AUTO range mode, the high and low ranges must be indepen-

dently calibrated. After pressing the CAL, CALZ or CALS keys, the user is prompted for the

range to be calibrated as seen in the zero calibration example below.