Teledyne 9110E - Nitrogen Oxides Analyzer User Manual
Page 64
Operating Instructions
Model 9110E Instruction Manual
50
M9110E Rev B
To enable or disable passwords, press the following keystroke sequence:
SAMPLE
RANGE = 500.0 PPB NO
x
=XXX.X
< TST TST > CAL
SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
SETUP X.X
PASSWORD ENABLE: OFF
OFF
ENTR
EXIT
SETUP X.X
PASSWORD ENABLE: ON
ON
ENTR
EXIT
SETUP X.X
PASSWORD ENABLE: ON
ON
ENTR EXIT
ENTR enables
the password
feature
Disable
or enable
passwords
Exit returns to
SAMPLE
display
Example: If all passwords are enabled, the following keypad sequence would be required to
enter the SETUP menu:
SAMPLE RANGE = 500.0 PPB NOX =XXX.X
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS: 0
0 0 0
ENTR EXIT
prompts for
password
number
Example: this
password enables the
SETUP mode
SAMPLE
ENTER SETUP PASS: 0
8 1 8
ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE
EXIT
Press individual
keys to set
numbers
Note that the instrument still prompts for a password when entering the VARS and DIAG
menus, even if passwords are disabled, but it displays the default password (818) upon
entering these menus. The user only has to press ENTR to access the password protected
menus but does not have to enter the required number code.