Preset Recall (:PRES:CALL)

This command recalls and inquires about preset data.

Setting

:PRES:CALL
is as shown below:
0: Cancels preset.
1: Calls PRESET 1 data.
2: Calls PRESET 2 data.
3: Calls PRESET 3 data.

Application
example

:PRES:CALL1
Recalls the set values stored in PRESET 1.

Query

:PRES:CALL?
Inquires about the setting.

Response
example

1
Indicates that PRESET 1 is selected. The query return
value has the same meaning as the setting .

Remark

This command causes an error if the sequence function is
running.

Preset Saving (:PRES:SAVE)

This command saves the current set values (voltage, current and
power) in PRESET memory.

Setting

:PRES:SAVE
is as shown below:
1: Saves the current set value in PRESET 1.
2: Saves the current set value in PRESET 2.
3: Saves the current set value in PRESET 3.

Application
example

:PRES:SAVE1
Saves the current set values in PRESET 1.

Query

None

Remark

This command causes an error if the sequence function is
running.